Face-mounted negative-pressure antechamber

A face-mounted chamber, developed during the COVID-19 pandemic, that contains aerosols created during endonasal surgery. Standard surgical suction pulls aerosol out of the chamber before it reaches the operating room.

U.S. Patent 11,490,988 B2

3D-printed negative pressure antechamber

ABOUT THE DEVICE

A barrier over the surgical field

Drilling bone in endoscopic endonasal surgery throws fine particles into the air. We sought to build a device that could capture aerosolized, potentially viral-bearing particles and would not get in the surgeon's way.

​

The antechamber sits on the patient's face and encloses the nose and mouth. A port on the front lets the surgeon pass the endoscope and instruments. A tube on top connects to standard hospital vacuum, which draws air and aerosol out of the chamber continuously.

Seals the face

​A soft silicone gasket seats the chamber against the patient's face. Three sizes cover different face shapes.

Pulls air out

​Suction on the top port keeps the inside at negative pressure, so aerosol is carried away instead of leaking out.

Work through the port

The surgeon operates through a silicone diaphragm that closes around the instruments. A version without the diaphragm gives more freedom of movement.

Evidence

The antechamber was tested in cadavers and used in 6 complex human cases, with airborne particle counts measured during bone drilling.

~97%

Fewer airborne particles (0.3–5 µm) during drilling in patients, compared with an uncovered face.

~60%

Fewer airborne particles (0.3–5 µm) during drilling in cadaver testing.

0

Complications in 6 complex human cases.

Reductions are calculated from the fold-increase in particle counts reported in the published studies (37-fold in patients, 2.49-fold in cadavers); with the antechamber, counts returned to baseline.

Published
Studies

2021

Development and validation of a patient face-mounted, negative-pressure antechamber for reducing exposure of healthcare workers to aerosolized particles during endonasal surgery

Lee, Rivera-Rosario, Kim, … Schwartz. Journal of Neurosurgery, 135:1825–1832.

2021

A novel negative pressure, face-mounted antechamber to minimize aerosolization of particles during endoscopic skull base surgery

Kim, Lee, Schwarz, Kacker, Schwartz. Operative Neurosurgery, 21:131–136.

COLLABORATE

Available for licensing

The invention is covered by U.S. Patent 11,490,988 B2.

Interested in licensing or collaboration?

Write to us if you want to build on this work, adapt it for another procedure, or discuss licensing. Licensing of the patent goes through Cornell, and we will point you to the right office.

Development and patent team

Development team

Authors of the Journal of Neurosurgery testing and validation study, from Weill Cornell Medicine and Cornell University.

Mark Lee

Matthew H. Kim

Jane Wang

Bradley Stylman

Aoife MacMahon

Theodore H. Schwartz

Hazel T. Rivera-Rosario

Gregory P. Bewley

Zellman Warhaft

Angela I. Park

Ashutosh Kacker

Patent

U.S. 11,490,988 B2. Assignee: Cornell University.

Ashutosh Kacker

Mark Lee

Theodore H. Schwartz

Availability
Team
Evidence

OtoMakers is an independent group and is not affiliated with, endorsed by, or speaking on behalf of any employer, hospital, or university. Notices specific to a device appear on that device’s project page.

© 2026 OtoMakers

bottom of page